IMPAC 测温仪
IGA 5

固定、全数字式、快速小型测温仪,配有模拟输出及RS232或RS485接口,可选装激光指示灯、目视瞄准系统或集成视频摄像机。测温范围250-2000 °C。

IMPAC 测温仪 IGA 5
IMPAC 测温仪 IGA 5

Key Features

  • excellent accuracy
  • long temperature ranges
  • laser pilot or view finder
  • analog and digital output
  • integrated maximum value storage
  • integrated TV camera
  • or PID controller (options)

Details

The pyrometers IS 5 and IGA 5 are digital, compact and fast infrared measuring instruments for non-contact temperature measurement on metals, ceramics or graphite.

For optimal match of the instrument to the application 2 different optics with extremely small spot sizes are available. The response time of only 2 ms facilitates the measurement of fast heating processes or short temperature peaks.

For a precise alignment of the pyrometers to the measuring object, the instruments are optionally equipped with a laser targeting light, a view finder or a video module. The most important parameters as emissivity, exposure time and analog output can be set directly in the instrument. Additionally the pyrometer can be connected to a PC via serial interface, enabling adjustments of further parameters with the delivered software InfraWin. This can be used for temperature indication, data logging and further analyzing of complete temperature processes.

Materials:

  • metal, steel
  • ceramics, graphite
  • glass, glass molds
  • alloys

Applications:

  • annealing
  • preheating
  • hardening, sintering
  • tempering, soldering
  • rolling
  • welding
  • normalising
  • shrinking, etc.

资料

Temperature range: (MB 18)    350 to 1800°C
(MB 20)    250 to 2000°C
(MB 25)    400 to 2500°C
(MB 30)    500 to 3000°C 
Sub range: User adjustable (minimum span is 51°C)
IR Detector: Indium-Gallium-Arsenic photo diode (InGaAs)
Spectral Range: 1.45 to 1.8 μm
Power Supply: : 24 V DC ± 25%, stabilized, ripple must be less than 50 mV
Power Consumption: ≤ 3 W (incl. active laser targeting light)
Analog Output: 0 - 20 mA or 4 - 20 mA, switchable, linear in temperature, load independent DC
Interface: Optional RS232 or RS485 (addressable), half duplex, baud rate 1.2 up to 38.4 kBd
Resolution: 0.1°C at the interface at the analog output < 0.1% of the adjusted temperature range but min. 0.1°C
Isolation: Power supply and digital output and analog output are galvanically isolated against each other
Parameters:

Adjustable on the converter’s rear side: emissivity slope, response time, analog output 0 to 20 mA or 4 to 20 mA, online / offline mode
Additionally via interface adjustable and readable: 2-color / 1-color temperature signal, according to this emissivity slope or emissivity, temperature sub range, settings for maximum value storage, address, baud rate, switch off limit, warning level lens contamination monitoring system.
Via interface readable only: measured value, internal temperature of the unit                    
Pyrometers with PID-controller via software adjustable: set point, proportional band, rate time/integral time, output delimitation.

Maximum Value Storage: Single or double storage, clear modes: time (off; 0.01 s; 0.05 s; 0.25 s; 1 s; 5 s; 25 s),
external clear contact, via interface or automatic „hot object mode“, hold-function for freezing the current temperature reading (not at pyrometers with PID-controller)
Emissivity ε: 0.2 to 1 adjustable in the instrument or with the software InfraWin in steps of 0.01
Response time t90: ≤ 2 ms, adjustable to 0.01 s; 0.05 s; 0.25 s; 1 s; 3 s; 10 s
Measurement uncertainty:
(Tamb.= 25°C, ε= 1, t90= 1 s)
 

< 350°C:             

0.5% of reading in °C + 1°C

350 to 1500°C:   

0.3% of reading in °C + 1°C

> 1500°C:           

0.5% of reading in °C + 1°C

Repeatability: 0.2% of reading in °C + 1°C (Tamb.= 25°C, ε= 1, t90= 1s)
Noise equivalent temperature difference (NETD):

0.4°C (t90= min=2 ms); 0.1°C (t90= 0.01 s)

(ε= 1, Tamb.= 10 to 40°C)

Sighting System: Laser targeting (max. power level < 1 mW, λ= 630-680 nm, CDRH class II)
or thru-lens view finder or video module
Ambient Temperature: 0 to 70°C
Storage Temperature: -20 to 70°C
Housing: Stainless steel, dimension see drawing on the right side
Weight: 550 g
Protection class: IP65 (DIN 40050)
CE-label: According to EU directives about electromagnetic immunity
   
Additional technical data for pyrometers with built-in video module:
Video signal: CCIR norm approx. 1 Vpp at 75 Ω, 50 Hz (special option: EIA norm 60 Hz, NTSC-M compatible)
Array size: CCIR norm: 628 x 583 pixels, black & white (EIA norm: 510 x 492 pixels, black & white)
Exposure: Automatic, additionally 3-levels controlled by the measuring temperature
Field of view: Approx. 10% x 14% of focused distance
Date/time: Real-time clock with at least 3 days spare run
Video output plug: Separate round plug at the pyrometer, not galvanically separated to the pyrometers power supply
Picture insertions: Target marking; unit number or user text (max. 12 characters), time and/or date; (individually switchable), measuring temperature, emissivity

 

IMPAC 测温仪 IGA 5
MB 18 & 25 Optics N a=90 mm
MB 20 Optics N a=90 mm
MB 18 & 25 Optics F a=220 mm
MB 20 Optics F a=220 mm
MB 18 & 25 Optics N a=150 mm
MB 18 & 25 Optics N a=250 mm
MB 20 Optik N a=150 mm
MB 20 Optics N a=250 mm
MB 18 & 25 Optics F a=500 mm
MB 18 & 25 Optics F a=1300 mm
MB 20 Optics F a=500 mm
MB 20 Optics F a=1300 mm
MB 18 & 25 Optics N a=100 mm
MB 18 & 25 Optics N a=200 mm
MB 18 & 25 Opitcs F a=300 mm
MB 18 & 25 Optics F a=800 mm
MB 18 & 25 Optics F a=2000 mm
MB 18 & 25 Optics F a=4000 mm
MB 20 Optics N a=100 mm
MB 20 Optics N a=200 mm
MB 20 Optics F a=300 mm
MB 20 Optics F a=800 mm
MB 20 Optics F a=2000 mm
MB 20 Optics F a=4000 mm
MB 30 Optics N a=90 mm
MB 30 Optics N a=100 mm
MB 30 Optics N a=150 mm
MB 30 Optics N a=200 mm
MB 30 Opitcs N a=250 mm
MB 30 Optik F a=220 mm
MB 30 Optics F a=300 mm
MB 30 Optics F a=500 mm
MB 30 Optics F a=800 mm
MB 30 Optics F a=1300 mm
MB 30 Optics F a=2000 mm
MB 30 Optics F a=4000 mm

光路显示:

查找您当地的代理

查找产品服务或买一台新的仪器.

全方位服务和支持

查找所有现场服务选项和资产保障.