校准源 M300

Mid Temperature Range Blackbody Calibration Source

Large calibration source with high emissivity for calibration independent of the wavelength. 200 ... 1150 °C (392° TO 2102°F)

Key Features

  • Superior accuracy within ±0.25% of reading ±1°C
  • Large 51 mm (2”) aperture diameter
  • High effective emissivity  1.0 at 0.7 to 1.8um, 0.998 typical at other wavelengths
  • Rugged housing designed for fixed installation in laboratory or test department
  • Manufactured and tested to meet rigid quality control standards
  • Furnished with certificate of calibration traceable to NIST
  • RS232 (standard) or RS485 (optional) serial communications port

Details

A high standard, precision blackbody calibration source designed with a unique uniformly heated spherical cavity that achieves a near ideal emissivity of 0.998 or better.

The M300 will deliver any temperature between 200 °C and 1150 °C (392° TO 2102 °F) – accurately. A convenient microprocessor-based, self-tuning digital PID controller holds the selected temperature to within 1 °C, thus assuring high accuracy calibration.

Typical Calibration Applications

  • Infrared Temperature Sensors
  • Infrared Thermal Imaging Systems
  • Spectrographic Analyzers
  • Spectral Radiometers
  • Heat Flux Meters

资料

Manual444 KB

M300 SPECIFICATIONS

Temperature Range: 200 °C to 1150 °C (392 °F to 2102 °F)
Temperature Uncertainty: ±0.25% of reading ±1 °C (when using Calibration Certificate correction factors)
Temperature Resolution: 0.1 °C
Stability: ±0.5 °C per 8 hour period
Source Non-Uniformity: ±0.1 % of reading  ±1 °C
Heated Cavity Shape: Spherical
Aperture Diameter: 51 mm (2”)
Emissivity: 1.0 effective at 0.65 to 1.8 um, 0.998 for others
Standard calibration method: Radiometric (pyrometric)
Temperature Sensor: Thermocouple
Remote Set Point: Via Serial Port
Method of Control: Digital Self tuning PID controller
Warm-up Time: approximately 1 hour from ambient to 1000 °C
Operating Ambient Temp: 0 °C to 44 °C (32 °F to 110 °F)
Cooling: Fan cooled, air inlet on rear panel
Operating Humidity: <90% RH non-condensing
Power Requirements: 208/230V AC ±10% 50/60 Hz 11A Maximum
Dimensions (H x W x D): 640 mm x 500 mm x 550 mm (25.2” x 19.7” x 21.7”) 
Weight: 80 kg (175 lbs.)
CE Certified: Yes

 

 