IMPAC 测温仪
IN 140/5
IN 140/5-L
IN 140/5-H
高精度、全数字式、快速测温仪,用于测量光谱响应为5.14 µm的玻璃表面温度。温度范围为250-2500 °C。
主要特点
- 测温范围250-2500 °C
- 响应时间极短,最小为10毫秒
- 光斑尺寸极小,最小为0.9毫米
- 调焦镜头
- 优化的目视取景器或激光靶光
- 内置数字式显示屏
- RS232 / RS485接口,可切换
- 测试电流输出
详情
IN 140/5是一款数字式测温仪,用于非接触式测量玻璃和石英玻璃表面的温度,温度范围250-2500 °C。该仪器的响应时间极短,光斑尺寸极小,适用于快速测量任务以及用于测量最小的测量物体。
IN 140/5机型配备了一个测头,具有更好的可视区域(2-NL光学镜头)。IN 140/5-H高速机型具有较短的响应时间,仅需10毫秒。为了使仪器适用于不同应用(测量物体的尺寸、测量距离),有不同的调焦镜头可供选择。激光靶光或优化的目视取景器可使仪器精确地对准测量物体。
测温仪配备了一个显示屏,在测量模式下可显示当前温度。此外,所有参数可通过仪器上的集成键盘进行更改并在显示屏上读取。
通过串行接口以及所配备的InfraWin软件,温度可以显示并存储在电脑内,还可设置参数。诊断功能能够检查仪器以及布线。
典型应用:
- 平板玻璃制造
- 灯泡生产
- 汽车玻璃装配
- 玻璃热弯
Technical Data |
|
Temperature ranges: |
MB 14 250 - 1400 °C MB 15 450 - 1500 °C MB 16 300 - 1600 °C MB 25 500 - 2500 °C |
Sub range: | any range adjustable within the temperature range, minimum span 51°C |
Data handling: | digital |
Spectral range: |
5.14 µm (narrow band, for glass surfaces) |
IR detector: | Thermopile |
Power supply: |
24 V AC or DC (14 to 30 V AC or DC) (AC: 48 to 62 Hz) |
Power consumption: | max. 1.2 W |
Analog output: | 0 - 20 mA or 4 - 20 mA (linear), switchable; test current 10 mA or 12 mA by pressing test key |
Load: | 0 to 500 Ω |
Digital Interface: | RS232 or RS485 addressable (half duplex), switchable; baud rate 1200 up to 115200 Bd |
Resolution:
|
0.1 °C on interface; 0.1 °C/°F (450 to 999 °C / 842 to 999 °F), 1°C (= 1000 °C/°F) on display; < 0.1% of the adjusted temperature sub range at the analog output |
Isolation: | power supply, analog output and digital interface are galvanically isolated from each other |
Operation signal: | green LED |
LC display: | illuminated LC display for temperature indication or parameter settings |
Parameters: | emissivity, exposure time, analog output, temperature sub range, settings of the maximum value storage, address, baud rate, internal temperature of the pyrometer. |
Emissivity ε: | 10 ... 120% adjustable in the instrument or via interface in steps of 0.1% |
Exposure time t90: |
IN 140/5; IN 140/5-L: 40 ms; adjustable up to 0.05 s; 0.25 s; 1 s; 3 s; 10 s IN 140/5-H: 10 ms; adjustable up to 0.05 s; 0.25 s; 1 s; 3 s; 10 s |
Maximum value storage: | built-in single or double storage. Clearing with adjusted time tclear (off; 0.01 s; 0.05 s; 0.25 s; 1 s; 5 s; 25 s), extern, via interface or automatically with the next measuring object |
Uncertainty: (ε = 1, t90 = 1 s) |
up to 1300 °C: 0.6% of reading in °C or 2 °C (Tamb. = 15 - 30 °C) *) 1% of reading in °C or 3 °C (Tamb. = 0 - 15 or 30 - 70 °C) *) above 1300 °C: 0.8 % of reading in °C (Tamb. = 15 - 30 °C) 1.2 % of reading in °C (Tamb. = 0 - 15 or 30 - 70 °C) *) whichever value is greater. The instrument must be at a constant ambient temperature for a minimum of 30 minutes |
Repeatability (ε = 1, t90 = 1 s): | 0.3% of measured value in °C |
Noise Equivalent Temperature Difference (NETD): (ε = 1, t90 = 1 s) Tamb.=10...40 °C) |
MB 15 / 25: at t90 = min: 1.2 °C (at 500 °C measuring temperature) at t90 = min: 0.6 °C (at 1200 °C measuring temperature) MB 14 / 16: at t90 = min: 0.7 °C (at 310 °C measuring temperature) at t90 = min: 0.2 °C (at 500 °C measuring temperature) at t90 = min: 0.15 °C (at 1200 °C measuring temperature) |
Sighting: | thru-lens view finder or laser targeting light (max. power level < 1 mW, l = 630-680 nm, CDRH class II) |
Ambient temperature: | 0 to 70°C |
Storage temperature: | -20 to 80°C |
Rel. humidity | non-condensing conditions |
Protection class: | IP65 (DIN 40050) |
Weight: | approx. 550 g |
CE-label: | according to EU directives about electromagnetic immunity |
IN 140/5 & 140/5-H | MB 14 & 16 | Optics 1-N |
IN 140/5 & 140/5-H | MB 14 & 16 | Optics 2-N |
IN 140/5 & 140/5-H | MB 14 & 16 | Optics 3-N |
IN 140/5 & 140/5-H | MB 15 & 25 | Optics 1-N |
IN 140/5 & 140/5-H | MB 15 & 25 | Optics 2-N |
IN 140/5 & 140/5-H | MB 15 & 25 | Optics 3-N |
IN 140/5-L | MB 14 & 16 | Optics 2-NL |
IN 140/5-L | MB 15 | Optics 2-NL |
IN 140/5-L | MB 15 | fixed optics a=163 mm |