校正源 M315X-HT (X4, X6, X8, X12)

Two-piece, high temperature blackbody calibration source with a large surface area.
Ambient + 5 … 600 °C (+ 9 ... 1112 °F)

校正源 M315X6-HT

Key Features

  • High emissivity
  • Excellent general purpose calibration
  • Large aperture sizes
  • High accuracy, high resolution
  • Manufactured and tested to meet rigid quality control standards 

Details

The M315X-HT blackbody calibration sources are designed to satisfy the exacting parameters of infrared focal plane array detectors, thermal imaging, and FLIR systems testing in projection scene and field applications where higher temperatures are required. These calibration sources combine high emissivity along with unchallenged stability and uniformity.

The M315X-HT blackbody sources are resistively heated by precision heating elements to provide uniform temperature distributions. The two-piece system is comprised of controller module and a separate enclosure for the emitter source with available aperture sizes from 4 x 4 inch (101 x 101 mm) up to 12 x 12 inch (305 x 305 mm), and temperature ranges from ambient +5.0° to 600.0 °C (+ 9 to 1112 °F). Emitter source temperature control is carried out by a precision digital PID controller, mounted in a 7" (180 mm) high 19" rack mount enclosure. An exceptionally stable RTD ensures long-term temperature stability of the system.

RS232C for remote control of set points and other critical parameters of the controller is standard. Optional RS485 output is available for remote operation.

Typical Calibration Applications

  • Infrared Focal Plane Detectors
  • Infrared Thermal Imaging Systems
  • FLIR systems testing

资料

Technical Specifications

Measurement Specifications
Temperature Range Ambient + 5 … 600 °C (+ 9 … 1112 °F)
Temperature Uncertainty1 ± 1 °C @ T < 100 °C … ± 2 °C @
T = 600 °C
Display Accuracy vs. NIST Calibration See supplied NIST calibration report
Temperature Resolution 0.1 °C
Stability2 ± 0.2 °C in still air environment
Source Non-Uniformity M315X4-HT: (in 3.5” x 3.5” region)
M315X6-HT: (in 4” x 4” region)
M315X8-HT: (in 6” x 6” region)
M315X12-HT: (in 10” x 10” region)
± 1.5 °C @ 200 °C
± 2 °C @ 400 °C
± 3 °C @ 500 °C
± 6 °C @ 600 °C
Heated Emitter Shape Flat Plate
Exit Port Diameter M315X4HT: 101 mm x 101 mm (4” x 4”)
M315X6HT: 152 mm x 152 mm (6” x 6”)
M315X8HT: 203 mm x 203 mm (8” x 8”)
M315X12HT: 305 mm x 305 mm (12” x 12”)
Emissivity ~ 1.00 effective emissivity @ 8 ... 14 μm
(A spectral emissivity graph is provided in the instruction manual)
Standard Calibration Method Radiometric
Temperature Sensor Precision platinum RTD
Warm-up Time < 30 minutes ambient ... 500 °C
Slew Rate to 1 °C Stability3 M315X4-HT: ~ 20°/min T < 300 °C
~ 10°/min T > 300 °C
M315X6-HT: ~30°/min T < 300 °C
~ 15°/min T > 300 °C
315X8-HT: ~ 20°/min T < 300 °C
~ 10°/min T > 300 °C
M315X12-HT: ~ 14°/min T < 300 °C
~ 7°/min T > 300 °C
Slew Rate to 0.1 °C Stability3 Approximately 30 minutes between
Δ 100 °C setpoints
Communication / Interface
Remote Set Point Via RS232 (standard) or RS485 (optional)
Method of Control Digital PID Controller
Environmental Specifications
Operating Ambient Temperature 10 °C ... 40 °C
Cooling Fan cooled, air inlet on rear panel
Operating Humidity < 90% non-condensing
Dimensions (H x W x D)

Blackbody Source:
M315X4-HT: 269 mm x 285 mm x 267 mm
(10.6” x 11.22” x 10.5”)
M315X6-HT: 417.6 mm x 406.4 mm x
371.4 mm (16.44” x 16” x 14.62”)
M315X8-HT: 417.6 mm x 406.4 mm
x 389.2 mm (16.44” x 16” x 15.32”)
M315X12-HT: 646.4 mm x 490.2 mm x 602 mm (25.45” x 19.3” x 23.7”)

Controller Dimensions:
195 mm x 432 mm x 576 mm (7.67" x 17" x 22.66")

Weight

Blackbody:
M315X4-HT: ~ 21 lbs (9.6 kg)
M315X6-HT: ~ 47 lbs (21.3 kg)
M315X8-HT: ~ 84 lbs (38.1 kg)
M315X12-HT: ~ 190 lbs (86.2 kg)

Controller: 3.2 kg (7 lbs)

CE Certified Yes
Electrical  
Power Requirements M315X4-HT – PN 19230-1:
220 ... 240 V AC @ 50 & 60 Hz 1050 VA
M315X4-HT – PN 19230-3:
115 V AC @ 50 & 60 Hz 1050 VA
M315X6-HT – PN 19100-4:
220 ... 240 V AC @ 50 & 60 Hz 4050 VA
M315X8-HT (EU) – PN 19200-5:
220 ... 240 V AC @ 50 & 60 Hz 4050 VA
M315X12-HT (EU & NA) – PN 18769-3:
220...240 V AC @ 50 & 60 Hz 9000 VA
M315X12-HT (NA) – PN 18769-4:
208 V AC @ 50 & 60 Hz 9000 VA

1 Accuracy calibration performed radiometrically, the uncertainty of emissivity and transfer standard are already included.
2 Provided stable AC mains voltage and minimum air flow across the exit port or emitter plate.
3 Typical. Can vary from unit to unit.