校准源 M390

Ultra-Fast High Temperature Blackbody Calibration Source

Calibration source for extremely high temperatures at very quick heat-up time. Two models cover 300° to 3000 °C (572° to 5432 °F)

校准源 M390 View

Key Features

  • Exceptional accuracy at high temperatures
  • ±0.25% of reading ± 1 °C
  • High effective emissivity of 1.0 between 0.65 to 1.8 um
  • Rapid warm up/cool down time – only 5 minutes from ambient to 2300 °C
  • Manufactured and tested to meet rigid safety and quality standards
  • Furnished with certification of calibration traceable to NIST
  • Remote calibration set point control via RS 232

Details

The M390 ultra-fast high temperature blackbody calibrator is without parallel in its capability to produce very high temperature, high emissivity targets, and at the same time stabilize at the required temperature within a few minutes.

The cavity is a graphite tube target with a 25 mm (1") diameter with an effective emissivity of 1.0 from 0.65 to 1.8um. Target temperature is sensed by a rapid response LumaSense infrared thermometer which drives a PID controller to regulate the target temperature precisely to the desired value.

The controller is mounted on the calibration source and remote set point programming may be achieved via the standard RS232 communication port. All safety features to protect equipment and operator are easily accessible.

Typical Calibration Applications

  • Infrared Temperature Sensors
  • Infrared Thermal Imagers
  • Spectroradiometers
  • High Energy Photon Generators
  • Heat Flux Gauges and Colorometers
  • Solar Radiance Simulation
  • Optical Pyrometer

资料

Size and Voltage Selections

 

M390-A2 – PN 14029-A2
Temperature Range 600-2300 °C  (1112-4172 °F)
Exit Aperture 25 mm (1.00")
Power 24kVA
M390-B2 – PN 14029-B2
Temperature Range   600-2600 °C  (1112-4712 °F)
Exit Aperture 25 mm (1.00")
Power 24kVA
M390-C2 – PN 14029-C2
Temperature Range 600-3000 °C  (1112-5432 °F)
Exit Aperture 25 mm (1.00")
Power 24kVA
M390-L1 – PN 14029-L1
Temperature Range 300-2000 °C  (572-3632 °F)
Exit Aperture 25 mm (1.00")
Power 15kVA

Technical Specifications

Accuracy*: ±0.25% of reading ±1 °C
Temperature Resolution: 1 °C
Stability: ±1 °C
Uniformity: Non-uniformity: ±0.1% ±1 °C of reading within the center ½ of diameter.
Cavity: Closed end graphite tube 292 mm (11.5") long with approximately 127 mm (5") heated length
Emissivity: 1.0 (effective from 0.6 to 1.8 um)
~0.96 (effective from 3 to 15 um)
Temperature Sensor: LumaSense Infrared Thermometer
Method of Control: Digital PID controller.
Remote Set Point: RS232 serial communication output standard, RS485 optional
Warm-up Time: 5 minutes from ambient to 2300 °C, typical
Operating Ambient Temp: 0 to 44 °C (32 to 110 °F)
Cooling: Water cooled. Standard (garden) hose connection on rear, 3.8 lpm (1 gpm minimum)
Cooling Inlet Pressure: 90 PSI (621 kPa) maximum. Customer is responsible to provide pressure safety valves at inlet and outlet.  Outlet (drain) must not be blocked.
Purge Gas: Argon High Purity (99.99 % pure)
Purge Gas Inlet Pressure: 25 PSI (172 kPa) maximum.  Customer is responsible to provide over pressure safety valves at inlet
Heating Element Type: Graphite tube with argon gas purge, 226 L/h  (8 to 10 cfh)
Operating Humidity: 0 to 90% RH non-condensing
Power Requirements: 208…230V AC @ 50 & 60 Hz, 15kVA or 24kVA (depends on model)
Power Connector: Customer Supplied
  • 15kVA  Must use plug that complies with IEC-309 at 75 Amps at 220/230V AC
  • 24kVA  Must use plug that complies with IEC-309 at125 Amps at 220/230V AC
Dimensions (H x W x D): 1710 mm x 560 mm x 820 mm (67.3” x 22” x 32.3”)
Weight: 182 kg (400 lbs.)
CE Certified: Yes

        *Accuracy calibration performed radiometrically, the uncertainty of emissivity and transfer standard are already included.