Kalibrierstrahler M305
Kompakte, universell einsetzbare Schwarzkörperkalibrierquelle für mittlere Temperaturen.
Temperaturbereich: 100 ... 1000 °C
Key Features
- Easily moved when needed. Weighs only 25 kg. (55 pounds)
- High accuracy ±0.2% of reading ±1 °C
- High emissivity 1.0 effective @ 0.7 ... 1.8 µm T>=230°C, 1.0 @ 8 ... 14 µm T<230°C
- Manufactured and tested to meet rigid quality control standards
- Furnished with certificate of calibration traceable to NIST
- RS232 (standard) or RS485 (optional) serial communications port
Details
The M305 blackbody calibration source uniquely combines portability with wide temperature range, high emissivity and high accuracy. It is an ideal calibration unit where portability is required or where installation space is limited. Its spherical, metal cavity design yields an emissivity of 0.995 over the temperature range of 100 to 1000 °C (212° to 1832 °F) with an aperture of 25 mm (1”). An integrally mounted PID controller adds to the unique versatility of this calibrator.
Quotations for custom designs and variations are available upon request.
Typical Calibration Applications
- Sapphire Probe Optical Pyrometer Calibrations
- Infrared Temperature Sensors
- Infrared Thermal Imaging Systems
- Spectrographic Analyzers
- Radiometers
- Flux Meters
Technical Specifications
Measurement Specifications | |
---|---|
Temperature Range: | 100 ... 1000 °C (210 ... 1832 °F) |
Temperature Uncertainty1: | ± 0.2% of reading ± 1 °C |
Temperature Resolution: | 0.1 °C |
Stability2: | 1 °C per 8 hour period |
Source Non-Uniformity: | ± 0.2 °C typical @ T < 230 °C, ± 1 °C typical at T > 230 °C |
Heated Cavity Shape: | Spherical |
Exit Port Diameter: | 25.4 mm (1.0”) |
Emissivity: | 0.995 ± 0.0005 (calculated from cavity shape) Effective Emissivity: 1.00 @ 8 ... 14 μm T < 230 °C, 1.00 @ 0.7 ... 1.8 μm T > 230 °C |
Standard calibration method: | Radiometric |
Temperature Sensor: | Thermocouple |
Warm-up Time: | 60 minutes from ambient to 700 °C |
Slew Rate to 1 °C Stability: | ~ 11 °C / min for Amb < T < 200 °C ~ 20 °C / min for Amb < T < 800 °C ~ 10 °C / min for T > 900 °C |
Slew Rate to 0.1 °C Stability: | 1 hour between setpoints |
Communication / Interface | |
---|---|
Remote Set Point: | Via serial port |
Method of Control: | Digital self tuning PID Controller |
Environmental Specifications | |
---|---|
Operating Ambient Temperature: | 0 ... 44 °C (32 ... 110 °F) |
Cooling: | Fan cooled, air inlet on rear panel |
Operating Humidity: | 90% RH max, non-condensing |
Dimensions (H x W x D): | 270 mm x 430 mm x 370 mm (10.6” x 16.9” x 14.6”) |
Weight: | 25 kg (55 lbs) |
CE Certified: | Yes |
Electrical | |
---|---|
Power Requirements: | 115 V AC @ 50 & 60 Hz or 230 V AC @ 50 & 60 Hz, 1000W |
1 Accuracy calibration performed radiometrically, the uncertainty of emissivity and transfer standard are already included.
2 Provided stable AC mains voltage and minimum air flow across the exit port or emitter plate.