IMPAC Pyrometer
IN 5/5

Compact pyrometer for temperature measurement of glass and quartz glass surfaces. Temperature ranges: 100 to 2500 °C

IMPAC Pyrometer IN 5/5 View
IMPAC Pyrometer IN 5/5 Back View Open

Key Features

  • Series IN 5/5: pyrometers in two wire form with analog output 4 to 20 mA, several temperature ranges available 
  • High accuracy due to digital linearisation of the output
  • Small spot sizes, min. 2.5 mm
  • Adjustable exposure time
  • Compact housing

Details

The pyrometer IN 5/5 is specially designed for non-contact temperature measurement of glass surfaces and quartz surfaces.

The IN 5/5 is a digital pyrometer in two wire technique. This technique combines the high accuracy of the digital signal processing with the simple connection and operating with two wires.

The solid and robust design of the instruments guarantees high operational safety even in rough industrial environments.

For optimal match of the instrument to the application (size of the measuring object, distance) different optics are available.

Typical Applications

  • Float glass
  • Hollow glass
  • Glass drop
  • Glass hardening
  • Glass bending
  • Bulb production
  • Heat treatment

Documentation

Temperature range:

(MB 6)   100 to  600°C 
(MB 8)   200 to  800°C
(MB 13) 100 to 1300°C
(MB 25) 400 to 2500°C
(Other MB on request)

Sub range: The plus instruments are user adjustable with minimum span of 51°C
Spectral Range: 5.14 μm
IR Detector: Thermopile
Data handling: Digital
Optics: Zinc-Sulfide lens (ZnS)
Power Supply:  24 V DC (10 to 30 V)
Power Consumption: max. 20 mA
Analog Output: 4 to 20 mA (linear)
Load: max. 700 Ω at 24 V (max. 100 Ω at 12 V)
Parameters Adjustable on the pyrometer: Emissivity, exposure time.
Emissivity ε: 0.2 ... 1 adjustable
Exposure Time t90: 0.08 s; adjustable in the pyrometer: 0.5 s; 1 s; 2 s; 5 s,

Measurement Uncertainty:

Dependent on object temperature T and
ambient temperature Tamb (ε= 1, t90= 1s)
 

T < 1300°C:

0.6% (IN 5/5-Lplus: 0.8%) of reading in °C or 2°C (Tamb=15 to 30°C) *)
1% of reading in °C or 1.5°C (Tamb=0 to 15 or 30 to 63°C) *)

T=1300 to 1800°C: 

0.8% of reading in °C (Tamb=15 to 30°C)
1.2% of reading in °C (Tamb=0 to 15 or 30 to 63°C)

T=1800 to 2500°C: 

1% of reading in °C (Tamb=15 to 30°C)
1.4 % of reading in °C (Tamb=0 to 15 or 30 to 63°C)

*) Whichever value is greater. The instrument must be at a constant ambient temperature for a minimum of 15 minutes (30 min for IN 5/5-L plus for 200 to 1300°C at Tamb = 0 to 15 or 30 to 63°C) and has to be connected to the power supply.

Repeatability: (ε = 1, t90= 1 s)
 

0.3% of reading in °C or 0.6°C
(Whichever value is greater. The instrument must be at a constant ambient temperature for a minimum of 15 min. (30 min for IN 5/5-Lplus for 200...1300°C at Tamb= 0 to 15 or 30 to 63°C)

Noise Equivalent Temperature Difference
(NETD): (ε = 1, Tamb= 23°C)
At t90= 80 ms: 0.7°C (at 110°C measuring temperature)
At t90= 1 s: 0.4°C (at 110°C measuring temperature)
Ambient Temperature: 0 to 70°C
Storage Temperature: -20 to 70°C
Housing: Stainless steel
Weight: 410g
Protection class: IP65 (DIN 40050)
Relative humidity: Non condensing conditions
CE-label: According to EU directives about electromagnetic immunity
IMPAC Pyrometer IN 5/5
MB 6 & 8 & 13: Optics 100 a=100 mm
MB 6 & 8 & 13: Optics 300 a=300 mm
MB 6 & 8 & 13: Optics 1200 a=1200 mm
MB 25: Optics 100 a=100 mm
MB 25: Optics 300 a=300 mm
MB 25: Optics 1200 a=1200 mm

Display of optical path:

Find Your Local Representative

To find a service for your product or to buy a new one.

Turn-key Service and Support

Find out all field services options and asset assurance.