Corps noir M315X (X4, X6, X8, X12)

Large Area Low Temperature Blackbody Source

Large area sources for calibrating thermal imaging systems, aerial mapping / surveillance equipment and spectrophotometers.
(Ambient +5) ... 400 °C

Corps noir M315X

Key Features

  • High emissivity and uniformity
  • Excellent general purpose calibration
  • Large aperture sizes
  • High accuracy, high resolution
  • Manufactured and tested to meet rigid quality control standards
  • RS232 (standard) or RS485 (option) serial communication output

Details

Designed to satisfy the exacting parameters of Infrared focal plane array Detectors, Thermal Imaging, and FLIR systems testing in projection scene and field application, the M315X Series blackbody calibration sources combine, high emissivity, and unchallenged stability and uniformity. Available aperture sizes ranges from 4 x 4 inch (101 x 101mm) up to 12 x 12 inch (305 x 305mm), and temperature ranges from ambient +5.0° to 400.0 °C.

The M315X Series Blackbody Sources are resistively heated by precision heating elements to provide uniform temperature distributions. The two-piece system is comprised of controller module and a separate enclosure for the emitter source.

Emitter source temperature control is carried out by a conventional, standard 19 inch rack mounting, precision digital PID controller. RS 232 is standard. RS485 is available as an option.  A standard connecting cable allows 12' (3.6m) of separation between source and controller. 

Typical Calibration Applications

  • thermal imaging systems,
  • aerial mapping / surveillance equipment
  • spectrophotometers

Documentation

Size and Voltage Selections

M315X4 – PN 19180-1
Temperature Range Ambient +5 to 400 °C
Exit Aperture 101 x 101 mm (4” x 4”)
Power 230V AC @ 50 & 60 Hz
M315X4 – PN 19180-4
Temperature Range Ambient +5 to 400 °C
Exit Aperture 101 x 101 mm (4” x 4”)
Power 115V AC @ 50 & 60 Hz
M315X6 – PN 19100-3
Temperature Range Ambient +5 to 400 °C
Exit Aperture 152 x 152 mm (6” x 6”)
Power 115V AC @ 50 & 60 Hz
No CE Certificate on these models
M315X8 – PN 19200-7
Temperature Range Ambient +5 to 400 °C
Exit Aperture 204 x 204 mm (8” x 8”)
Power 115V AC @ 50 & 60 Hz
M315X12 – PN 19080-5
Temperature Range Ambient +5 to 400 °C
Exit Aperture 305 x 305 mm (12” x 12”)
Power 230V AC @ 50 & 60 Hz

Technical Specifications

Temperature Uncertainty
(Thermometric calibration):
0.01% or 150 mK, whichever is greater
Temperature Uncertainty
(Radiometric calibration):
±1 °C at T < 100 °C to ± 1.3 °C at T = 400 °C
Display Accuracy vs. NIST Calibration: See separate calibration report
Temperature Resolution: 0.01°
Short term stability: ±0.1°C per 8 hours period if in still air environment (X4, X6 versions)
±0.2 °C if in still air environment (X8, X12 versions)
Source Non-Uniformity: ±0.4 °C (within 3.5” x 3.5” region) at 100 °C (X4 version)
±0.4 °C @ 100°C within center 4” x 4” region (X6 version)
±0.6 °C over 6” x 6” region @ 100 °C (X6 version)
±1 °C (within 6” x 6” region) at 100 °C (X8 version)
±1 °C (within 10” x 10” region) at 100 °C (X12 version)
Heated Emitter Shape: Flat Plate
Emissivity
(standard radiometric calibration):
effective @ 8-14 µm: 1.00 (Spectral Emissivity graph is provided in the instruction manual)
Emissivity
(optional thermometric calibration):
@ 8-14 µm: 0.96
@ 3-5 µm: 0.936
Slew Rate to 1 °C Stability: <60 minutes maximum to 400 °C (X4, X6 versions)
30 minutes maximum to 200 °C, <1 hr to 400 °C (X8, X12 versions)
Temperature Sensor: Precision platinum RTD 1/3 DIN
Remote Set Point: Via RS 232 (standard) or RS 485 (optional)
Method of Control: Digital PID controller
Operating Ambient Temp: 15 °C to 40 °C (X4, X6 versions)
15 °C to 40 °C (X8, X12 versions)
Operating Humidity: <90% non condensing
Dimensions (H x W x D): X4 version: 280 mm x 254 mm x 280 (11” x 10” x 11”)
to X12 version: 510 mm x 660 mm x 585 (20” x 26” x 23”)
Controller module: 178 mm x 483 mm x 593 mm (7”h x 19”w x 23”d)
Weight (Controller): Approximately 2 lbs (9kg)
Weight (Blackbody): Approximately 315x4: 12 lbs (5.4kg), 315x12: 163 lbs (74kg)