Черное тело Модель M330-EU

High Temperature Blackbody Calibration Source

High temperature calibration source, 300 °C to 1700 °C (572 °F to 3092 °F)

Calibration source M330 View
Calibration source M330-EU View

Key Features

  • High effective emissivity 1.0 at 0.65-1.8 um
  • High accuracy
  • Excellent stability +/-1°C per 8 hour period
  • Fast slew rate (limited to 20°C/min)
  • Manufactured and tested to meet rigid quality control standards
  • Furnished with certificate of calibration traceable to NIST
  • Wide temperature range
  • RS 232 serial communication (standard), RS 485 (optional)

Details

The M330 is a blackbody calibration source utilizing a digital indicating temperature controller that may be set to any temperature between 300 °C (572 °F) and 1700 °C (3092 °F). A precision thermocouple controls the blackbody cavity providing high accuracy and repeatability. A closed-end tube with a 25 mm (1”) aperture diameter is heated by specially manufactured elements.  The temperature controller uses the industry standard PID algorithms to control the emitter temperature to within +/-0.5 °C. The blackbody uses a resistive heater that provide long life. (Running at temperatures greater than 1600 °C for long periods will shorten heater life.)  An independent over-temperature power cutout system is an added safety feature.  An internal fan keeps the cabinet surfaces at a safe, comfortable temperature.

Typical Calibration Applications

  • Infrared Temperature Sensors
  • Infrared Thermal Imagers
  • Spectroradiometers
  • High Energy Photon Generators
  • Solar Radiance Simulation
  • Optical Pyrometers 

документация

M330-US / M330-EU SPECIFICATIONS

Temperature Range: 300 °C to 1700 °C (572 °F to 3092 °F)
Accuracy*: ±0.25% of reading ±1 °C
Temperature Resolution: 0.1 °C
Stability: ±1 °C per 8 hour period
Source Non-Uniformity ±1 °C within center 1/3 of ID
Cavity: Closed end tube 41 mm ID X 255 mm long with ≈125 mm heated length.
Aperture Diameter: 25 mm (1”)
Emissivity: 1.0 at 0.65-1.8 um (lower at longer wavelengths-a correction table is supplied in the manual)
Calibration Method: Radiometric (pyrometric)
Temperature Sensor: Type B thermocouple
Remote Set Point: Via serial port
Method of Control: Digital Self-Tuning PID Controller
Warm-up Time: 45 min. from ambient to 300 °C
65 to 80 min. from 300 to 1600 °C
Operating Ambient Temp: 0 °C to 44 °C (32 °F to 110 °F)
Cooling: Fan cooled, air inlet on rear panel
Operating Humidity: <90% RH, non- condensing
Power Requirements:

208...240V AC, 50 & 60 Hz, (US), 3000VA Maximum

230V AC, 50 Hz (EU)
Dimensions (H x W x D):

648 mm (H) x 500 mm (W) x 551 mm (D) (25.52” x 19.7” x 21.7”) 

Weight: 80 kg (175 lbs.)
CE Certified: Yes

        *Accuracy calibration performed radiometrically, the uncertainty of emissivity and transfer standard are already included

Найти местных представителей

To find a service for your product or to buy a new one.

Turn-key Service and Support